Medtech

Testing for a leak in a medical implant

The Challenge

Creating a test to verify that a medical implant could last 10 years in the body without leaking, in order to pass regulatory controls and ultimately launch it to market.

 

Our Approach

We developed a three stage test to inspect for leaks. Stage 1 was a visual inspection. Stage 2 used the rapid helium bombing method to catch large leaks and finally stage 3 involved a long helium bombing process whereby any small leaks were detected.

We tested the protocol with known leaky samples and edge cases to ensure that the test method produced valid results across a representative range of parts. Once we had confidence in the test method we proceeded to design verification and validation, and subsequently our clients could launch the product.

 

The Result

Developing verification tests can be more complex than first envisioned. It is important to have a detailed technical understanding of the requirements and product to ensure that verification methods and acceptance criteria reliably prove the safety and effectiveness of the product. For novel designs, it can be necessary to think creatively about the most valuable methods of verification and to create bespoke test protocols.

Gilero News

Gilero, a Sanner Group Company, starts manufacturing in Greensboro, USA

Life at Gilero

Completing a Graduate Degree while Working at Gilero

Manufacturing

The Importance of Current Good Manufacturing Practices (cGMPs) in Medical Device Manufacturing

General

Heuristic Analysis: An Alternative to Formative Usability Testing  

Gilero News

Sanner Group Acquires Gilero to Expand its Global Medical Device Offering

Drug Delivery

Developing a Simpler Nasal Vaccine Administration Device for Infants

Drug Delivery

Beyond Borders: How geopolitical shifts are redefining pharma and medical device development